CONTACTLESS AND CONTACT-BASED SMARTPHONE PHOTOPLETHYSMOGRAPHY AS A MULTI-PARAMETER CARDIOMETABOLIC HOME SCREEN: BEYOND ATRIAL FIBRILLATION TOWARD BLOOD PRESSURE AND DYSGLYCEMIA ESTIMATION

Authors

DOI:

https://doi.org/10.31435/ijitss.3(51).2026.6042

Keywords:

Photoplethysmography, Smartphone Health, Atrial Fibrillation Screening, Cuffless Blood Pressure, Digital Biomarkers, Remote Monitoring

Abstract

Smartphone photoplethysmography (PPG) demonstrates a stark maturity gradient across cardiometabolic parameters. For atrial fibrillation (AF) detection, meta-analytic evidence from 14 studies (n = 5,090) shows pooled sensitivity of 96% (95% CI 93 to 97%) and specificity of 97% (95% CI 95 to 98%), with individual validation studies reporting sensitivity up to 99.7% and specificity up to 99.7% using machine learning algorithms in ambulatory settings. AF detection performance is consistent across multiple smartphone devices, robust in unsupervised home use, and clinically impactful: population screening increased oral anticoagulation uptake from 56% to 74%, and digital post-ablation monitoring detected arrhythmia recurrence at more than double the rate of conventional ECG follow-up (38.5% versus 17.7%, OR 3.4). Signal quality rates for PPG (89 to 97%) were comparable to or exceeded those for single-lead ECG. Blood pressure (BP) estimation, by contrast, remains unreliable for population-level screening. Of 18 studies evaluated, only two clearly passed established AAMI/ESH/ISO validation protocols, both relying on per-individual calibration against a reference measurement. The two largest prospective trials (n = 965 and n = 62) explicitly failed validation standards, and contactless remote PPG showed critically low sensitivity for hypertension detection (systolic BP sensitivity 0.04) in a diverse-skinned field population. Diastolic BP was consistently estimated more accurately than systolic BP across approaches. Dysglycemia detection rests on a single large study demonstrating moderate screening discrimination (AUC 0.74 to 0.77) using a deep neural network applied to PPG waveforms, with the score independently associated with HbA1c. The evidence supports smartphone PPG as a clinically ready tool for AF detection and monitoring, particularly in populations aged over 65, but BP estimation requires individual calibration and cannot yet substitute for cuff-based measurement, and dysglycemia screening remains investigational.

References

Avram, R., Olgin, J., Kuhar, P., et al. (2020). A digital biomarker of diabetes from smartphone-based vascular signals. Nature Medicine, 26(10), 1576–1582. https://doi.org/10.1038/s41591-020-1010-5

Barbosa, I. O. F., Oliveira, B. C. de, Santos, C. K. M., et al. (2025). Smartphone-based applications for atrial fibrillation detection: A systematic review and meta-analysis of diagnostic test accuracy. Telemedicine Journal and e-Health. https://doi.org/10.1089/tmj.2024.0579

Brasier, N., Raichle, C. J., Dörr, M., et al. (2018). Detection of atrial fibrillation with a smartphone camera: First prospective, international, two-centre, clinical validation study (DETECT AF PRO). Europace, 21(1), 41–47. https://doi.org/10.1093/europace/euy176

Calvert, P., Mills, M. T., Howarth, K., et al. (2024). Remote rhythm monitoring using a photoplethysmography smartphone application after cardioversion for atrial fibrillation. European Heart Journal – Digital Health. https://doi.org/10.1093/ehjdh/ztae028

Chan, P., Wong, C.-K., Poh, Y. C., et al. (2016). Diagnostic performance of a smartphone-based photoplethysmographic application for atrial fibrillation screening in a primary care setting. Journal of the American Heart Association, 5(7), e003428. https://doi.org/10.1161/JAHA.116.003428

Dasa, D., & Davies, P. (2025). Equity and feasibility of remote photoplethysmography for hypertension screening in darker-skinned populations in Nigeria: A multi-site field study. medRxiv. https://doi.org/10.1101/2025.10.13.25337878

Degott, J., Ghajarzadeh-Wurzner, A., Hofmann, G., et al. (2021). Smartphone based blood pressure measurement: Accuracy of the OptiBP mobile application according to the AAMI/ESH/ISO universal validation protocol. Blood Pressure Monitoring, 26(6), 441–448. https://doi.org/10.1097/MBP.0000000000000556

Delmotte, L., Desebbe, O., Alexander, B., et al. (2023). Smartphone-based versus non-invasive automatic oscillometric brachial cuff blood pressure measurements: A prospective method comparison volunteer study. Journal of Personalized Medicine, 14(1), 15. https://doi.org/10.3390/jpm14010015

Dörr, M., Weber, S., Birkemeyer, R., et al. (2018). iPhone app compared with standard RR measurement: The iPARR trial. SSRN. https://doi.org/10.2139/ssrn.3266975

Dörr, M., Weber, S., Birkemeyer, R., et al. (2020). iPhone app compared with standard blood pressure measurement: The iPARR trial. American Heart Journal, 233, 102–108. https://doi.org/10.1016/j.ahj.2020.12.003

Eckstein, J., Burkardt, T., Winterhalder, C., et al. (2017). Use of a smartphone camera to estimate systolic blood pressure compared to standard oscillometric blood pressure measurement: A prospective trial. European Heart Journal, 38(Suppl. 1). https://doi.org/10.1093/eurheartj/ehx501.P441

Fan, Y.-Y., Li, Y., Li, J., et al. (2019). Diagnostic performance of a smart device with photoplethysmography technology for atrial fibrillation detection: Pilot study (Pre-mAFA II Registry). JMIR mHealth and uHealth, 7(3), e11437. https://doi.org/10.2196/11437

Fernstad, J., Svennberg, E., Åberg, P., et al. (2022). Validation of a novel smartphone-based photoplethysmographic method for ambulatory heart rhythm diagnostics. EP Europace, 24(Suppl. 1). https://doi.org/10.1093/europace/euac053.578

Fernstad, J., Svennberg, E., Åberg, P., et al. (2024). Validation of a novel smartphone-based photoplethysmographic method for ambulatory heart rhythm diagnostics: The SMARTBEATS study. Europace, 26(5), euae079. https://doi.org/10.1093/europace/euae079

Fernstad, J., Svennberg, E., Åberg, P., et al. (2025). External validation of a machine learning-based classification algorithm for ambulatory heart rhythm diagnostics in pericardioversion atrial fibrillation patients using smartphone photoplethysmography: The SMARTBEATS-ALGO study. Europace, 27(2), euaf031. https://doi.org/10.1093/europace/euaf031

Freithaler, M., Chandrasekhar, A., Dhamotharan, V., et al. (2023). Smartphone-based blood pressure monitoring via the oscillometric finger-pressing method: Analysis of oscillation width variations can improve diastolic pressure computation. IEEE Transactions on Biomedical Engineering, 70(11), 3052–3063. https://doi.org/10.1109/TBME.2023.3275031

Gill, S., Sartini, C., Uh, H., et al. (2020). Accurate detection of atrial fibrillation using a smartphone remains uncertain: A systematic review and meta-analysis. European Heart Journal, 41(Suppl. 2). https://doi.org/10.1093/ehjci/ehaa946.3505

Gogiberidze, N., Sagirova, Z., Kuznetsova, N., et al. (2021). Comparison of cuffless blood pressure measurement using an electrocardiogram monitor with photoplethysmography function with measurement by the Korotkoff method: A pilot study. Sechenov Medical Journal, 12(1), 39–49. https://doi.org/10.47093/2218-7332.2021.12.1.39-49

Gruwez, H., Evens, S., Proesmans, T., et al. (2021). Evaluation of the device-independent nature of a photoplethysmography-deriving smartphone app. Europace, 23(Suppl. 3). https://doi.org/10.1093/europace/euab116.525

Gruwez, H., Evens, S., Proesmans, T., et al. (2021). Head-to-head comparison of proprietary PPG and single-lead ECG algorithms for atrial fibrillation detection. EP Europace, 23(Suppl. 3). https://doi.org/10.1093/europace/euab116.524

Gruwez, H., Verbrugge, F. H., Proesmans, T., et al. (2023). Smartphone-based atrial fibrillation screening in the general population: Feasibility and impact on medical treatment. European Heart Journal – Digital Health, 4(6), 464–472. https://doi.org/10.1093/ehjdh/ztad054

Gruwez, H., Ezzat, D., Van Puyvelde, T., et al. (2024). Real-world validation of smartphone-based photoplethysmography for rate and rhythm monitoring in atrial fibrillation. Europace, 26(5), euae065. https://doi.org/10.1093/europace/euae065

Gruwez, H., Van Herendael, H., Maris, S., et al. (2024). Effectiveness of photoplethysmography-based rhythm monitoring after atrial fibrillation ablation using a smartphone application: The DIGITOTAL study. Heart Rhythm, 22(4). https://doi.org/10.1016/j.hrthm.2024.11.026

Gruwez, H., Melio, N. D., Vermunicht, P., et al. (2025). Improving atrial fibrillation or flutter detection and management by smartphone-based photoplethysmography rhythm monitoring following cardiac surgery: A pragmatic randomized trial. Europace, 27(2), euaf015. https://doi.org/10.1093/europace/euaf015

Guo, Y., Wang, H., Zhang, H., et al. (2019). Mobile health technology for atrial fibrillation screening using photoplethysmography-based smart devices: The HUAWEI Heart study. Journal of the American College of Cardiology, 74(19), 2365–2375. https://doi.org/10.1016/j.jacc.2019.08.019

Hereijgers, M., Van der Velden, R. M. J., Reinders, L., et al. (2025). Feasibility and diagnostic yield of a photoplethysmography-based arrhythmia screening and integrated management pathway in a COPD outpatient clinic. ERJ Open Research, 11(1). https://doi.org/10.1183/23120541.01192-2024

Holyoke, P., Yogaratnam, K., & Kalles, E. (2021). Web-based smartphone algorithm for calculating blood pressure from photoplethysmography remotely in a general adult population: Validation study. Journal of Medical Internet Research, 23(1), e19187. https://doi.org/10.2196/19187

Khan, M., Evens, S., Childe, S., et al. (2023). Remote heart rhythm assessment using smartphone-based photoplethysmography in primary care to detect undiagnosed atrial fibrillation in patients with chronic obstructive pulmonary disease. European Heart Journal, 44(Suppl. 2). https://doi.org/10.1093/eurheartj/ehad655.2981

Konečný, T., & Saxon, L. (2016). Integration of smartphone-based monitoring: No modality is an island, entire of itself. Journal of Cardiovascular Electrophysiology, 27(1). https://doi.org/10.1111/jce.12844

Lamberigts, M., Van Hoof, L., Proesmans, T., et al. (2021). Remote heart rhythm monitoring by photoplethysmography-based smartphone technology after cardiac surgery: Prospective observational study. JMIR mHealth and uHealth, 9(8), e26519. https://doi.org/10.2196/26519

Lawin, D., Hoffmann, A., Lawrenz, T., et al. (2026). Prolonged smartphone-based photoplethysmography for heart rhythm monitoring after repeat ablation of atrial fibrillation: A 6-month prospective study on mHealth compliance, motivation and arrhythmia recurrence. Heart, Lung and Circulation. https://doi.org/10.1016/j.hlc.2025.08.028

Maitas, O., Lee, J., Robotis, D. A., et al. (2012). Detection of atrial fibrillation using an iPhone 4S. Circulation, 126(Suppl. 21), A9298. https://doi.org/10.1161/circ.126.suppl_21.a9298

Manninger, M., Hermans, A., Caracioni, A., et al. (2023). Photoplethysmography-documented atrial fibrillation in the first week after catheter ablation is associated with lower success rates. Frontiers in Cardiovascular Medicine, 10, 1199630. https://doi.org/10.3389/fcvm.2023.1199630

Mena, L., Félix, V., Ostos, R., et al. (2020). Mobile personal health care system for noninvasive, pervasive, and continuous blood pressure monitoring: Development and usability study. JMIR mHealth and uHealth, 8(7), e18012. https://doi.org/10.2196/18012

Mortelmans, C., Proesmans, T., Van Haelst, R., et al. (2019). Smartphone-based use of the photoplethysmography technique to detect atrial fibrillation in primary care: A diagnostic accuracy study of the FibriCheck app.

Mutke, M. R., Brasier, N., Raichle, C. J., et al. (2020). Comparison and combination of single-lead ECG and photoplethysmography algorithms for wearable-based atrial fibrillation screening. Telemedicine Journal and e-Health, 26(12), 1440–1446. https://doi.org/10.1089/tmj.2020.0036

Napolitano, C., Chong, J., Saczynski, J., et al. (2015). Pulse-based arrhythmia discrimination using a novel smartphone application. Circulation, 132(Suppl. 3), A11586. https://doi.org/10.1161/circ.132.suppl_3.11586

Oladunni, T., & Adewumi, F. G. (2026a). Attractor-vascular coupling theory: Formal grounding and empirical validation for AAMI-standard cuffless blood pressure estimation from smartphone photoplethysmography.

Oladunni, T., & Adewumi, F. G. (2026b). Cardiac stability theory: An axiomatically grounded framework for continuous cardiac health monitoring via smartphone photoplethysmography.

Proesmans, T., Mortelmans, C., Van Haelst, R., et al. (2018). Mobile phone-based use of the photoplethysmography technique to detect atrial fibrillation in primary care: Diagnostic accuracy study of the FibriCheck app [Preprint]. https://doi.org/10.2196/preprints.12284

Proesmans, T., Mortelmans, C., Van Haelst, R., et al. (2019). Mobile phone-based use of the photoplethysmography technique to detect atrial fibrillation in primary care: Diagnostic accuracy study of the FibriCheck app. JMIR mHealth and uHealth, 7(3), e12284. https://doi.org/10.2196/12284

Proesmans, T., Smeets, C., Dreesen, P., et al. (2019). Preliminary results of the FLASH-AF: Validation of the device-independent nature of a pulse-deriving smartphone application for the detection of atrial fibrillation. European Heart Journal, 40(Suppl. 1). https://doi.org/10.1093/eurheartj/ehz748.0222

Proesmans, T., Vanacker, P., Verbrugge, F. H., et al. (2020). Digital-AF II: Impact of smartphone-based atrial fibrillation screening in the general population for primary stroke prevention. Stroke, 51(Suppl. 1), A134. https://doi.org/10.1161/str.51.suppl_1.134

van Putten, L. D., Bamford, K. E., Veleslavov, I., & Wegerif, S. (2024). From video to vital signs: Using personal device cameras to measure pulse rate and predict blood pressure using explainable AI. Discover Applied Sciences, 6, 218. https://doi.org/10.1007/s42452-024-05848-8

Qiao, D., Ayesha, A. H., Zulkernine, F., et al. (2022). ReViSe: Remote vital signs measurement using smartphone camera. IEEE Access, 10, 131656–131670. https://doi.org/10.1109/ACCESS.2022.3229977

Rajan, S., Sivapuram, M., Kumar, S., & Podder, V. (2023). Accuracy of smartphone-based vital monitoring using remote photoplethysmography technology enabled WellFie application. medRxiv. https://doi.org/10.1101/2023.01.14.23284548

Reichl, J., Hamid, K. A., Burkard, T., et al. (2025). PPG-based smartphone application versus usual care for atrial fibrillation screening: A European multicenter randomized trial. Heart Rhythm, 22(11). https://doi.org/10.1016/j.hrthm.2025.07.060

Rozen, G., Vaid, J., et al. (2018). Diagnostic accuracy of a novel mobile phone application for the detection and monitoring of atrial fibrillation. American Journal of Cardiology, 121(10), 1187–1191. https://doi.org/10.1016/j.amjcard.2018.01.035

Sagirova, Z., Kuznetsova, N., Gogiberidze, N., et al. (2021). Cuffless blood pressure measurement using a smartphone-case based ECG monitor with photoplethysmography in hypertensive patients. Sensors, 21(10), 3525. https://doi.org/10.3390/s21103525

Sollee, J., Cheema, B., Slotwiner, D., et al. (2025). FibriCheck detection capabilities for atrial fibrillation: A multicenter validation study. Circulation, 152(Suppl. 3), 4338116. https://doi.org/10.1161/circ.152.suppl_3.4338116

Soto, J., Hershman, S., & Ashley, E. (2021). Combining digital data and artificial intelligence for cardiovascular health. Cardiovascular Research, 117(11), 2274–2276. https://doi.org/10.1093/cvr/cvab211

Szonyi, M. D., Gausz, F., Bocz, B., et al. (2026). Diagnostic performance of two commercially available, PPG-based smartphone applications to detect atrial fibrillation. European Heart Journal – Digital Health, 7. https://doi.org/10.1093/ehjdh/ztag009

Tabei, F., Gresham, J. M., Askarian, B., et al. (2020). Cuff-less blood pressure monitoring system using smartphones. IEEE Access, 8, 11534–11545. https://doi.org/10.1109/ACCESS.2020.2965082

Tan, S. Y., Chai, J., Choi, M., et al. (2024). Remote photoplethysmography technology for blood pressure and hemoglobin level assessment in the preoperative assessment setting: Algorithm development study. JMIR Formative Research, 8, e60455. https://doi.org/10.2196/60455

Vischer, A., Rosania, J., Socrates, T., et al. (2022). Comparability of a blood-pressure-monitoring smartphone application with conventional measurements: A pilot study. Diagnostics, 12(3), 749. https://doi.org/10.3390/diagnostics12030749

Yamakoshi, T., Rolfe, P., & Yamakoshi, K. (2020). Cuffless blood pressure estimation based on haemodynamic principles: Progress towards mobile healthcare. PeerJ, 8. https://doi.org/10.7717/peerj.11479

Downloads

Published

2026-08-27

How to Cite

Buszko, J., Dyś, M., Buż, A., Danilczuk, D., Bajkowska-Piterak, M., Szlachta-Gubernat, M., Baran, N., Piterak, P., & Adamiec, W. (2026). CONTACTLESS AND CONTACT-BASED SMARTPHONE PHOTOPLETHYSMOGRAPHY AS A MULTI-PARAMETER CARDIOMETABOLIC HOME SCREEN: BEYOND ATRIAL FIBRILLATION TOWARD BLOOD PRESSURE AND DYSGLYCEMIA ESTIMATION. International Journal of Innovative Technologies in Social Science, 3(3(51). https://doi.org/10.31435/ijitss.3(51).2026.6042

Most read articles by the same author(s)

<< < 1 2